Reference Executive Standard
GB/T10151-2008 Medical diagnostic X-ray equipment-Specifications for high voltage cable plugs and socketsProduct Features
1、Mini receptacle,for CA11 high voltage plug 2、Small size 75kVDC,100℃ rated receptacle 3、Thermoplastic material with high flame retardant UL 94-5VA and high insulation resistance(≥1015Ω ·m) 4、Optional brass thrust ring 5、Optional O-shaped rubber ring for oil seal 6、Optional nickel plated brass flange 7、Cathode socket---Four wires 8、Anode socket---Single wire or all four wires are connected to each otherProduct picture
![details](http://kehu02.grofrom.com/www.sailray-tube.com/fa8777344f703e05e3aa21b1720b5d4e.jpg)
![certificate](http://kehu02.grofrom.com/www.sailray-tube.com/21fe26c7d3a0f6bf1161b1cb6e2ad909.jpg)
Technical data
Rated voltage between pin and flange | 75kVDC |
Number of contact pins | 4 |
Maximum continuous rated voltage between pins | 4kVDC |
Maximum continuous current (per pin) | 15A |
Insulation resistance between pin and flange | >1015Ω |
Maximum continuous operating temperature | 100℃ |
Product Drawing